Check category
/
/
/
Maintenance of ultrasonic diagnostic equipment

Maintenance of ultrasonic diagnostic equipment

  • Categories:Company news
  • Author:
  • Origin:
  • Time of issue:2022-01-19 17:31
  • Views:

(Summary description)1. Operating environment and site requirements for ultrasonic diagnostic equipment

Maintenance of ultrasonic diagnostic equipment

(Summary description)1. Operating environment and site requirements for ultrasonic diagnostic equipment

  • Categories:Company news
  • Author:
  • Origin:
  • Time of issue:2022-01-19 17:31
  • Views:

1. Operating environment and site requirements for ultrasonic diagnostic equipment

    (1) Pressure: 70~106KPa;

    (2) Examination room: no direct sunlight, good air convection. There is no interference from power distribution devices and X-ray devices around, and it is far away from high magnetic field environments;

    (3) After confirming that the power switch on the instrument and the voltage stabilizer is in the "OFF" position, insert the plug into the power socket.

    (4) Whether the power supply voltage is stable and whether it is used together with other electrical devices will affect the performance of the ultrasonic diagnostic apparatus.

    (5) After the machine is turned on, pay attention to whether the mechanical sound emitted by the instrument is normal, and observe whether the self-checking procedure is correct;

    (6) Once there is a sudden power failure, the power supply of the instrument should be restored to the "OFF" position immediately; after the voltage is stable, restart the machine;

    (7) If the instrument suddenly stops automatically, the power supply should be cut off, and professional maintenance personnel are required to help;

    (8) First, turn off various operations and set the control keys to the state at the beginning of use;

    (9) After the instrument is fully dissipated, cover it with the instrument cover.

 

 2. Ultrasound system maintenance and repair

    (1) Regularly clean all exterior surfaces of the instrument with a milder detergent and a damp cloth.

    (2) The air filtration device of the system should be inspected every week, up to 2 weeks, and cleaned if necessary.

    (3) Turn off the power switch when removing the filter device, and do not start the host system when the air filter plate is not installed.

 

Third, the maintenance and maintenance of the probe

   1. Do not immerse the entire probe in water or other liquids. Below the cable connection of the probe is not waterproof.

   2. After the probe is used, wipe the ultrasonic coupling agent clean. When cleaning the probe, use a mild detergent and a damp cloth. The probe should be kept clean frequently.

Keyword:

Recommended

Category brief summary: 新闻资讯
Maintenance of ultrasonic diagnostic equipment
Maintenance of ultrasonic diagnostic equipment
Time of issue : 2022-01-19 17:31:54
1. Operating environment and site requirements for ultrasonic diagnostic equipment
See more information
1. Operating environment and site requirements for ultrasonic diagnostic equipment
How helpful are fundus cameras for fundus screening?
How helpful are fundus cameras for fundus screening?
Time of issue : 2022-01-19 17:29:32
Compared with the method of examining fundus lesions using fundoscopy, fundus cameras have the advantages of being intuitive, accurate, and easy to use, and can provide more valuable reference for ophthalmologists.
See more information
Compared with the method of examining fundus lesions using fundoscopy, fundus cameras have the advantages of being intuitive, accurate, and easy to use, and can provide more valuable reference for ophthalmologists.

Add:NO.124 Sijia industrial zone 2 Hengjie Taizhou city

Zhejiang province China


Mob:13634023343
Tel:0576-82788886 
Fax:0576-82626863
Email:
weizhen@tz-wz.com

Copyright © 2022 Zhejiang Weizhen Optical Instrument Technology Co.,Ltd. 浙ICP备2022003076号 Powered by 300.cn