这里是标题一h1占位文字
Maintenance and Care of Ultrasound Diagnostic Equipment
Release time:
2022-01-18 17:18
1. Requirements for the operating environment and location of the ultrasound diagnostic device
(1) Pressure: 70~106KPa;
(2) Examination room: No direct sunlight, good air circulation. No interference from power distribution devices, X-ray devices, etc., and away from high magnetic field environments;
(3) Confirm that the power switch on the instrument and stabilizer is in the "OFF" position before plugging the power plug into the socket.
(4) Whether the power supply voltage is stable and whether it is used together with other electrical devices will affect the performance of the ultrasound diagnostic device.
(5) After turning on the machine, pay attention to whether the mechanical sound emitted by the instrument is normal, and observe whether the self-check program is correct;
(6) In the event of a sudden power outage, immediately restore the instrument's power to the "OFF" position; wait for the voltage to stabilize before restarting;
(7) If the instrument suddenly stops automatically, disconnect the power and seek help from professional maintenance personnel;
(8) First, set all operation shutdown and control keys to the state they were in when first used;
(9) After the instrument has fully dissipated heat, cover it with the instrument cover.
2. Maintenance and repair of the ultrasound system
(1) Regularly clean all external surfaces of the instrument with a mild detergent and a damp cloth.
(2) The system's air filtration device should be checked weekly, or at most every two weeks, and cleaned if necessary.
(3) When removing the filtration device, turn off the power switch, and do not start the main system when the air filter plate is not installed.
3. Maintenance and upkeep of the probe
1. Never immerse the entire probe in water or other liquids. The cable connection part of the probe is not waterproof.
2. After using the probe, wipe off the ultrasound coupling agent. When cleaning the probe, use a mild detergent and a damp cloth. The probe should be kept clean regularly.
Related News
Understanding the Importance and Functionality of the Slit Lamp in Ophthalmic Examinations
The slit lamp is a versatile and indispensable tool in the field of ophthalmology, allowing healthcare professionals to conduct detailed examinations of the anterior segment of the eye. This instrument combines a high-intensity light source with a biomicroscope, enabling doctors to observe the eye's structures in a magnified view. Its design allows for both a broad observation and a more focused e
Why LCD Vision Charts Enhance Diagnostic Accuracy for Eye Health
Why LCD Vision Charts Enhance Diagnostic Accuracy for Eye Health Introduction to LCD Vision Charts In an era where technology permeates every facet of healthcare, the adoption of LCD vision charts marks a significant advancement in ophthalmic diagnostics. These innovative tools are not just replacements for traditional printed charts; they represent a paradigm shift in how vision assessments are c
How a Slit Lamp Enhances Diagnostic Accuracy for Eye Conditions
How a Slit Lamp Enhances Diagnostic Accuracy for Eye Conditions Introduction to Slit Lamps The slit lamp is a cornerstone of modern ophthalmic practice, providing a magnified view of the anterior segment of the eye, including the cornea, iris, and lens. This instrument combines a high-intensity light source with a microscope, allowing healthcare professionals to assess the eye's structures in deta
Enhancing Visual Assessment: The Role of LCD Vision Charts in Medical Diagnostics
In the realm of medical diagnostics, the ability to effectively assess a patient's vision is paramount. One of the most innovative tools available to healthcare providers is the LCD Vision Chart. This device has revolutionized the way vision tests are conducted, providing numerous benefits over traditional methods. LCD Vision Charts utilize advanced digital technology to present a wide range of vi
Why Every Ophthalmologist Should Consider a Digital Slit Lamp: A Comprehensive Guide
Why Every Ophthalmologist Should Consider a Digital Slit Lamp Introduction: The Evolution of Ophthalmic Equipment In today's fast-paced medical environment, ophthalmologists are faced with the pressure of providing accurate diagnoses and effective treatments while maintaining high patient satisfaction. With technological advancements, the traditional slit lamp has evolved into the **digital slit l
Understanding Auto Refractometers: A Key Tool in Modern Ophthalmology
Auto refractometers are vital instruments widely used in ophthalmology and optometry to assess a patient’s refractive error. They provide a quick and accurate measurement of how light is focused in the eye, helping eye care professionals determine the appropriate corrective lenses for various vision problems such as myopia, hyperopia, and astigmatism. The operation of an auto refractometer is rel